- LoCut : 120 W @ 100 Ohms CF 1.5
- HiCut : 120 W @ 100 Ohms CF 1.5
- ScalCut : 200 W @ 100 Ohms (5 Levels) CF 1.5
Bipolar Coagulation Modes:
- Cut/Standard : 120 W @ 100 Ohms CF 1.5
- Macro/Force : 120 W @ 100 Ohms CF 1.5
- Micro/Precise : 100 W @ 100 Ohms CF 1.5
- Auto Micro: Bipolar Micro with Auto-Stop
- Auto Start: Auto-Start, Auto-Stop
Bipolar Saline Resection:
- LowCut : 250 W @ 200 Ohms CF 1.5
- HiCut 1: 300 W @ 200 Ohms CF 1.5
- HiCut 2: 300 W @ 200 Ohms CF 1.5
Bipolar Saline Coagulation:
- Desi : 250 W @ 25 Ohms CF 1.5
- Force: 250 W @ 25 Ohms CF 2.1
- Input : 88-265V AC, 50/60 Hz
- Power Consumption : 750VA Max
- Input Fuse : 5A T
- Output Control : Colour TFT Display with Touch Screen and Push buttons in all Modes
- Max. Power Output : 400 Watts
Safety Standard
- Basic Construction : As per IEC 60601-1, IEC 60601-1-2 & IEC 60601-2-2
- Protection Class : I
- Mdd. Class : II b
- Pollution degree Class : II
- Overvoltage category Class : II
- Unit Type : CF
- All Outputs : Floating Output (All output ports are CF Type)
- Electrical Potential Balancing : Connecting Pin and Indicated by Symbol
- PER : 98
- Dimension : 335mm (W) x 160mm (H) x 395mm (D)
- Weight: Maximum 8.0 Kg
- Ambient temperature range: +10 C to +40 C
- Relative humidity: 30% to 75% non-condensing
- Atmospheric pressure: 700 millibars to 1060 millibars
- Ambient temperature range : -30 C to +65 C
- Relative humidity: 0% to 90% non-condensing
- Atmospheric pressure: 500 millibars to 1060 millibars