- Pure Monopolar : Cut mode is for Underwater (TURP) and General Surgeries
- Low Cut : Mode specifically tailored for Laparoscopy
- Blend: Cutting with Coagulation effect. MONOPOLAR
COAGULATION
- Fulgurate/ Force A low spray waveform suitable for non-contact coagulation oozing surfaces and (TURP) underwater coagulation
- Desiccate : Standard Coagulation Mode.
BIPOLAR
- Cut/Standard: Best coagulation mode when used with graspers. especially designed for Bipolar Scissors.
- Force/Macro: Faster Bipolar Coagulation
RF output characteristics
MONOPOLAR CUT
- Pure : 300 Watts/300 Ohms
- Low Cut : 200 Watts/300 Ohms
- Blend : 100 Watts/300 Ohms
- MONOPOLAR COAGULATION
- Fulgurate : 100 Watts/500 Ohms
- Desiccate : 100 Watts/500 Ohms
BIPOLAR
- Cut/Standard : 80 Watts / 100 Ohms
- Force/Macro : 80 Watts/100 Ohms
INPUT SUPPLY
- Input Supply : 230VAC + 20% 50 Hz
- Power Consumption : 500VA Max.
- Loading Duty Factor 10Sec (Active)/ 30Sec (Pause)
- Input Fuse : 3.15AT
- Output Control : Digital Display, Push buttons in all Modes.
SAFETY STANDARDS
Basic Construction: Accordance with
- IEC 60601-1,
- IEC 60601-1-2 &
- IEC 60601-2-2
- Protection Class : 1
- Unit Type: CF
- All Outputs: Floating Outputs
- Electrical Potential: Connecting Pin and
- Balancing: Indicated by Symbol.
- Weight: 3.2 Kg.
- Size (in mm): 310(W) x 150(H) x 280(D)